If you've ever used an oscilloscope to look at a square wave in electronics, or did any audio synthesis, you look at the ending of the post and raise an eyebrow (in the way Spock does it), and note that yes, this is what square waves are in real life. The corner wobble is normal.
I recall a research paper claiming to have measured a finite amount of time for an electron to jump between atomic energy levels. If true, then it is likely everything takes at least some time. Except if you yourself move at light speed.
Except that no driver has infinite bandwidth, and no connection has perfect impedance along the entire length from start to finish. The "wobble" will reappear when you a scope with an even higher sample rate. Which means it is real.
Different signals have different rise times. In fact in any meaningful high frequency device (above say 10 MHz) the electrical engineers layouting the PCBs in your device must know those rise times, otherwise they are building a transmitting antenna.
It is pretty simple: the faster a transition is the higher its bandwidth. A theoretical instant transition would require infinite frequency bandwidth. In reality anything conducting electricity is fast, but not faster than the speed of light and every transmission line will have real world characteristics that limit its bandwidth at some point.
So the rise time and ringing of signals is not just an artifact of the oscilloscope (although it can be if your probes are not calibrated). Overshoot and ringing can really be there.
Well it can be. But you will also encounter squares that don't overshoot at all, but rather undershoot.
Also, you probably know that your oscilloscope probes have a small trimmer inside them that is to be adjusted to get rid of exactly this artifact. You hook up the probe to the calibrate output of your scope and turn that trimmer till the square wave from the calibrate output is indeed square.
If your scope always showed that ringing, your probe is broken or needs calibration.
Replies (4)
Replies (3)
Replies (3)
It is pretty simple: the faster a transition is the higher its bandwidth. A theoretical instant transition would require infinite frequency bandwidth. In reality anything conducting electricity is fast, but not faster than the speed of light and every transmission line will have real world characteristics that limit its bandwidth at some point.
So the rise time and ringing of signals is not just an artifact of the oscilloscope (although it can be if your probes are not calibrated). Overshoot and ringing can really be there.
Also, you probably know that your oscilloscope probes have a small trimmer inside them that is to be adjusted to get rid of exactly this artifact. You hook up the probe to the calibrate output of your scope and turn that trimmer till the square wave from the calibrate output is indeed square.
If your scope always showed that ringing, your probe is broken or needs calibration.